Technology

Instrument Sharing

    1、Terahertz Spectrometer

        Brand Model:TeraPulse4000

图片7.png

        Specifications: Dynamic range >50dB, Frequency range up to 4THz, Scanning speed > 5Hz, with independent complete transmission, reflection, ATR measurement module

        Charge: 400 RMB/ hour


    2、 Gas Chromatography Mass Spectrometry

        Brand Model: Agilent:GC-MS 7820A-5977E

图片8.png 

 

(一)Mass Spectrometry:
    1、Mass Spectrometry Performance 

        1)、Fully automatic tuning
        2)、The number of mass range: 1.6-1050amu, 0.1amu increments
        3)、Resolution: single mass count
        4)、The quality axis stability: not less than 0.10amu / 48 hours
        5)、sensitivity:
            Full scan sensitivity (electron bombardment source): 1pg OFN, the signal / noise ratio> 400 ︰ 1 

            Selection detection (electron bombardment source): 20fg OFN, the signal / noise ratio> 10 ︰ 1
        6)、Maximum scan rate: 12500u / s 1 ms SIM dwell time
        7)、Dynamic range: full dynamic range of 106
        8)、Selection detection mode: desirable 100 groups of ions, 60 ions / group
        9)、With traction detection function
    2、Ion Source 

        1)、Made of completely inert material (non-coated, non-stainless steel)
        2)、EI source with two long-term filament, can be automatically switched by the software
        3)、Ionization energy: 5-241.5eV
        4)、Ion source temperature: independent temperature control, 150-350˚C adjustable
    3、The Quality Analyzer: 

        1)、The whole quartz gold-plated hyperboloid quadrupole
        2)、Specially designed Turner-Kruge lens, can improve the sensitivity, protect the quadrupole,,
        3)、Can be independent temperature control, the maximum temperature can be ≧ 200˚C
    4、The Detector: long-term high-energy electronic multiplier
    5、The Vacuum System: high-performance molecular turbo pump


(二)Chromatographic:

    1. Column oven

        1)Size: 28.0 × 30.5 × 16.5 cm
        2)Temperature range: room temperature above 8 ℃-425℃
        3)Temperature setting accuracy: 1℃
        4)The fastest heating rate: 75 ℃/min
            50-70°C, heating rate: 75°C/min
            70-115°C, heating rate: 45°C/min
            115-175°C, heating rate: 40°C/min
            175-300°C, heating rate: 30°C/min
            300-425°C, heating rate: 20°C/min
        5)The longest running time: 999.99 minutes
        6)Program temperature: 5 order
        7)Temperature accuracy: 1 ° C per room temperature, temperature fluctuation of<0.01 ° C
        8)oven temperature program Accuracy: ≤ 2%:≤2%
        9)Program temperature reproducibility: ≤1%:≤1%

    2.Heating Area
        1)Independent heating zone, excluding the column oven: 5 (two inlets, another detector, one auxiliary)
        2)Auxiliary area maximum operating temperature: 350 ℃
        3)Supports up to two heating valves


    3.Electronic airway control (EPC)
        1)EPC control is possible for all inlets and detectors
        2) Injection module::
           &nbsp  Pressure control accuracy: 0.01 psi psi
             Flow sensor accuracy:< ± 5%
              Detector Module Accuracy: 8%

    4.Inlet:
        1)Up to two inlets can be installed
        2)Mountable: packed column inlet and split / splitless inlet


    5.Fill column inlet::
        1) Electronic flow control
        2)Maximum operating temperature 400 ° C
        3)Maximum flow:< 100 mL/min
        4)To accommodate 1/4 inch and 1/8 inch packed columns and 0.530 mm capillary columns

    6.Split/splitless inlet (S/SL)
        1)Electronic pressure / flow control
        2)Maximum operating temperature 400 ° C
        3)Pressure range: 0 to 60 psi
        4)Maximum split ratio: 250: 1
        5)Total flow setting range: N2: 0 ~ 200 mL / min; H2 or He: 0 ~ 1250 mL / min;H2或He:0~1250 mL/min

 

3、 Near Infrared Spectrometer

图片9.png 

Brand Model: Ocean Optical Flame-NIR

Introduction to parameters: 

Detector response range: 900-1700 nm

Wavelength range: 950-1650 nm (preconfigured model)

Optical resolution: ~ 10 nm FWHM (25μm)

Signal to noise ratio: 6000: 1 (full signal)

Dynamic range: 6000: 1 (single acquisition)

Thermal stability: 0.08 nm / ° C

Scanning frequency (max): ~ 400Hz

Integration time: 1 ms → 65 s

Pixels: 128

Size: 89.1 mm x 63.3 mm x 31.9 mm

(34.4 mm with pad)

Weight: 265 g

 

4、 Microscope

图片10.png 

Brand Model: LEICA DM4M:LEICA DM4M

Parameters:

Metallographic microscope

Modular design, can achieve reflection observation, transparent reflection observation configuration

Achromatic optical path, the overall optical path supports 25mm field diameter

6-hole objective lens turntable, with 32mm diameter industrial objective lens

Observation mode can be realized bright field, dark field, polarized light, interference

 

5、 Autocorrelaton Function Analyzer

图片11.png 

Brand Model: APE Company Pulsecheck USB50

Parameters:

PulseCheck USB 50/measurable pulse width:<50Fs-12PS